    ULTRAMAT 23, 19" rack unit, dimensions in mm

    ULTRAMAT 23, bench-top unit, dimensions in mm

    Схема подключения

    Pin assignment (electrical and gas connections)

    ULTRAMAT 23, pin assignment (standard)

    ULTRAMAT 23, pin assignment of optional PROFIBUS interface card

    19" rack unit

    ULTRAMAT 23, 19" rack unit, e.g. an IR measured component with oxygen measurement

    Portable unit

    ULTRAMAT 23 portable, in sheet-steel enclosure, gas connections and electrical connections

    ULTRAMAT 23, designation of the different labels

    Технические данные

    19" rack unit and portable version

    General information

     

    Measured components

    Maximum of 4, comprising up to 3 infrared-sensitive gases and oxygen

    Measuring ranges

    2 per measured component

    Display

    LCD with LED backlighting and contrast control; function keys; 80 characters (4 lines/20 characters)

    Operating position

    Front wall, vertical

    Conformity

    CE symbol EN 61000-6-2, EN 61000-6-4

    Design, enclosure

     

    Weight

    Approximately 10 kg

    Degree of protection, 19" rack unit and desktop model

    IP20 according to EN 60529

    Electrical characteristics

     

    EMC interference immunity (electromagnetic compatibility)
    Safety extra-low voltage (SELV) with safe isolation

    In accordance with standard requirements of NAMUR NE21 (08/98) or EN 50081-1, EN 50082-2

    Auxiliary power

    100 V AC, +10%/-15%, 50 Hz,

    120 V AC, +10%/-15%, 50 Hz,

    200 V AC, +10%/-15%, 50 Hz,

    230 V AC, +10%/-15%, 50 Hz,

    100 V AC, +10%/-15%, 60 Hz,

    120 V AC, +10%/-15%, 60 Hz,

    230 V AC, +10%/-15%, 60 Hz

    Power consumption

    Approx. 60 VA

    Electrical inputs and outputs

     

    Analog output

    Per component, 0/2/4 up to 20 mA, NAMUR, isolated, max. load 750 Ω

    Relay outputs

    8, with changeover contacts, freely parameterizable, e.g. for measuring range identification;

    24 V AC/DC/1 A load, potential-free, non-sparking

    Digital inputs

    3, dimensioned for 24 V, potential-free

    • Pump
    • AUTOCAL
    • Synchronization

    Serial interface

    RS 485

    AUTOCAL function

    Automatic unit calibration with ambient air (depending on measured component),
    adjustable cycle time from 0 (1) … 24 hours

    Options

    Add-on electronics, each with 8 additional digital inputs and relay outputs for e.g. triggering of automatic calibration and for PROFIBUS PA or PROFIBUS DP

    Climatic conditions

     

    Permissible ambient temperature

     
    • During operation

    5 to 45 °C

    • During storage and transportation

    -20 ... +60 °C

    Permissible ambient humidity

    < 90% RH (relative humidity) during storage and transportation

    Permissible pressure fluctuations

    600 … 1 200 hPa

    Gas inlet conditions

     

    Sample gas pressure

     
    • Without pump

    Unpressurized (< 1 200 hPa, absolute)

    • With pump

    Depressurized suction mode, set in factory with 2 m hose at sample gas outlet; full-scale value calibration necessary under different venting conditions
    (800 ... 1 050 hPa, absolute)

    Sample gas flow

    72 … 120 l/h (1.2 … 2 l/min)

    Sample gas temperature

    Min. 0 to max. 50 °C, but above the dew point

    Sample gas humidity

    < 90% RH (relative humidity), non-condensing



    Infrared channel

    So that the technical data can be complied with, a cycle time of ≤ 24 hours must be activated for the AUTOCAL. The cycle time of the AUTOCAL function must be ≤ 6 hours when measuring small NO and SO2 measuring ranges (≤ 400 mg/m³) on TÜV/QAL-certified systems.

    Measuring ranges

    See ordering data

    Chopper compartment flushing

    Upstream pressure approximately 3 000 hPa; purging gas consumption approximately 100 ml/min

    Time response

     

    Warm-up period

    Approximately 30 min (at room temperature) (the technical specification will be met after 2 hours)

    Delayed display (T90 time)

    Dependent on length of analyzer chamber, sample gas line and parameterizable attenuation

    Damping (electrical time constant)

    Parameterizable from 0 … 99.9 s

    Measuring response

    Based on sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature

    Output signal fluctuation

    < ± 1% of the current measuring range (see rating plate)

    Detection limit

    1% of the current measuring range

    Linearity error

    • In the largest possible measuring range:
      < ± 1% of the measuring range full scale value
    • In the smallest possible measuring range:
      < ± 2% of the measuring range full scale value

    Repeatability

    ≤ ± 1% of the current measuring range

    Drift

     

    Zero point

     
    • With AUTOCAL

    Negligible

    • Without AUTOCAL

    < 2 % of the current measuring range/week

    Full-scale value drift

     
    • With AUTOCAL

    Negligible

    • Without AUTOCAL

    < 2 % of the current measuring range/week

    Influencing variables

    Based on sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature

    Temperature

    Max. 2% of the smallest possible measuring range according to rating plate per 10 K with an AUTOCAL cycle time of 6 h

    Atmospheric pressure

    < 0.2% of the current measuring range per 1% pressure variation

    Auxiliary power

    < 0.1% of the current measuring range with a change of ± 10%



    Oxygen channel (electrochemical)

    Measuring ranges

    0 … 5 % … 0 … 25 % O2, parameterizable

    Service life

    Approx. 2 years with 21 % O2

    Detection limit

    1% of the current measuring range

    Time response

     

    Delayed display (T90 time)

    Dependent on dead time and parameterizable attenuation, not > 30 s at approximately 1.2 l/min sample gas flow

    Measuring response

    Based on sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature

    Output signal fluctuation

    < ± 0.5% of the current measuring range

    Linearity error

    < ± 0.2% of the current measuring range

    Repeatability

    ≤ 0.05 % O2

    Drift

     
    • With AUTOCAL

    Negligible

    • Without AUTOCAL

    1% O2/year in air, typical

    Influencing variables

    Based on sample gas pressure 1 013 hPa absolute, 1.0 l/min sample gas flow and 25 °C ambient temperature

    Temperature

    < ± 0.5% O2 per 20 K, relating to a measured value at 20 °C

    Atmospheric pressure

    < 0.2% of the measured value per 1% pressure variation

    Accompanying gases

    The oxygen sensor must not be used if the accompanying gas contains the following components: Chlorine or fluorine compounds, heavy metals, aerosols, mercaptans, alkaline components (such as NH3 in % range)

    Typical combustion exhaust gases

    Influence: < 0.05% O2

    Humidity

    H2O dew point ≥ 2 °C; the oxygen sensor must not be used with dry sample gases (however, no condensation either)



    H2S channel for measuring ranges of 5 ... 50 vpm

    Measured components

    Maximum of 4, comprising up to 2 infrared-sensitive gases, an oxygen component and a hydrogen sulfide component

    Measuring ranges

     
    • Smallest measuring range

    0 ... 5 vpm

    • Largest measuring range

    0 ... 50 vpm

    Service life of the sensor

    Approx. 12 months

    Permissible atmospheric pressure

    750 ... 1 200 hPa

    Permissible operating temperature

    5 ... 40 °C (41 ... 104 °F)

    Operating mode

    • Continuous measurement between 0 and 12.5 vpm
    • Discontinuous measurement between 12.5 and 50 vpm

    Influencing variables

     

    Accompanying gases

    The hydrogen sulfide sensor must not be used if the accompanying gas contains the following components:

    • Compounds containing chlorine
    • Compounds containing fluorine
    • Heavy metals
    • Aerosols
    • Alkaline components
      (e.g. NH3 > 5%)

    Cross-inferences (interfering gases)

    1 360 vpm SO2 result in a cross-interference of < 20 vpm H2S

    180 vpm NO result in a cross-interference of < 150 vpm H2S

    No cross-interference of CH4, CO2 and H2 (1 000 vpm)

    Drift

    < 1% of the current measuring range per month

    Temperature

    < 3% /10 K referred to full-scale value

    Atmospheric pressure

    < 0.2% of the measured value per 1% pressure variation

    Measuring response

    Delayed display (T90 time)

    < 40 s with sample gas flow of approx. 1 ... 1.2 l/min

    Output signal noise

    < 2% of smallest measuring range with an attenuation constant of 30 s

    Display resolution

    < 0.01 vpm H2S

    Output signal resolution

    < 1% of smallest measuring range with an attenuation constant of 30 s

    Repeatability

    < 4% of smallest measuring range



    Paramagnetic oxygen cell

    Measured components

    Maximum of 4, comprising up to 3 infrared-sensitive gases and an oxygen component

    Measuring ranges

    2 per component

    • Min. 0 ... 2% vol O2
    • Max. 0 ... 100% vol O2
    • Suppressed measuring range possible; e.g. 95 ... 100%

    Permissible atmospheric pressure

    700 ... 1 200 hPa

    Permissible operating temperature

    5 ... 45 °C (41 ... 113 °F)

    Cross-inferences (interfering gases)

    See Table "Cross-sensitivities"

    Zero point drift

    • Measuring range 2%: max. 0.1% with weekly zero adjustment
    • Measuring range 5%: max. 0.1% with weekly zero adjustment
    • Measuring range 25% or greater:
      max. 0.5% with monthly zero adjustment

    Measured-value drift

    Negligible with AUTOCAL

    Temperature error

    < 2%/10 K referred to measuring range 5%

    < 5%/10 K referred to measuring range 2%

    Humidity error for N2 with 90% relative humidity after 30 min

    < 0.6% at 50 °C

    Atmospheric pressure

    < 0.2% of measured value per 1% pressure variation

    Delayed display (T90 time)

    < 60 s

    Output signal noise

    < 1% of smallest measuring range

    Repeatability

    < 1% of smallest measuring range



    Дальнейшая информация

    Ordering notes

    Special selection rules must be observed when measuring some components.

    Measured component N2O

    7MB2335, 7MB2337 and 7MB2338
    (application: Si chip production)

    • Measuring range 0 to 100 / 500 vpm (MB designation "E")
    • Can only be used to measure N2O in ultra-pure gases

    7MB2337 and 7MB2338
    (application: measurement in accordance with the requirements of the Kyoto protocol)

    • Measuring range 0 to 500 / 5 000 vpm (MB designation "Y")
    • Requires simultaneous measurement of CO2 for correction of cross-interference

    7MB2337-*CP*0-*SY* or
    7MB2338-*DC*0-*SY* (including NO measurement)

    7MB2337 and 7MB2338
    (application with paramagnetic oxygen measuring cell and separate gas path)

    7MB2337-4**80-**** - Z + C11
    7MB2337-5**80-**** - Z + C11

    7MB2338-4**80-**** - Z + C11
    7MB2338-5**80-**** - Z + C11

    Measured component SF6

    7MB2335, 7MB2337 and 7MB2338
    (application: SI chip production)

    • Measuring range 0 to 500 / 2 500 vpm (MB designation "H")
    • Can only be used to measure SF6 in inert gases

    Calibration interval (MCERTs versions 7MB2335, 7MB2337, 7MB2338)

    Component

    Smallest measuring range

    Calibration interval

    Remarks

    Z suffix

    CO

    0 to 150 mg/m³

    5 months

    IED 2010/75/EC

    E50

    CO

    0 … 250 mg/m³

    12 months

    IED 2010/75/EC

     

    NO

    0 … 100 mg/m³

    5 months

    IED 2010/75/EC

     

    NO

    0 … 250 mg/m³

    12 months

    IED 2010/75/EC

     

    SO2

    0 to 400 mg/m³

    12 months

    IED 2010/75/EC

     

    N2O

    0 … 500 vpm

     

    Kyoto protocol

     

    N2O

    0 … 50 mg/m³

    6 months

    IED 2010/75/EC

     


    Calibration interval (TÜV versions 7MB2355, 7MB2357, 7MB2358)

    Component

    Smallest measuring range

    Calibration interval

    Remarks

    Z suffix

    CO

    0 ... 200 mg/m³

    1 month

    13th/27th BImSchV

    T13/T23/T33

    NO

    0 to 150 mg/m³

    1 month

    13th/27th BImSchV

    T13/T23/T33

    SO2

    0 to 400 mg/m³

    1 month

    13th/27th BImSchV

    T13/T23/T33



     

    AUTOCAL
    (ambient air)

    AUTOCAL
    (inert gas e.g. N2)

     

    Calibration with calibration gas

    Comment (keep to technical specs)

    Zero point

    Calibration point

    Zero point

    Calibration point

    Zero point

    Calibration point

    Hours

    Weeks

      

    IR components

    3 ... 24

    3 ... 24

    o

    52

     

    O2 - electrical chemical sensor

    Stable

    3 ... 24

    Stable

    -

    52

    o

     

    O2 paramagnetic Cell

    -

    3 ... 24

    x

    x

    1

    o

    at MB < 5%

    -

    3 ... 24

    x

    x

    8

    o

    at MB > 5%

    O2 paramagnetic Cell

    x

    x

    3 ... 24

    -

    o

    52

    at MB < 5%

    x

    x

    3 ... 24

    -

    o

    52

    at MB > 5%

    H2S sensor

    3

    -

    3

    -

    o

    4

     

    o = with AUTOCAL, x = not applicable



    Calibration intervals, standard devices